Field Metrology Wells

- Lightweight, portable, and fast
- Cool to –25 °C in 15 minutes and heat to 660 °C in 15 minutes
- Built-in two-channel readout for PRT, RTD, thermocouple,
4-20 mA current
- True reference thermometry with accuracy to ± 0.01 °C
- On-board automation and documentation
- Metrology performance in accuracy, stability, uniformity, and loading
|
|
|
|
9142
|
9143
|
9144
|
|
Temperature Range at 23 °C
|
–25 °C to 150 °C (–13 °F to 302 °F)
|
33 °C to 350 °C (91 °F to 662 °F)
|
50 °C to 660 °C (122 °F to 1220 °F)
|
|
Display Accuracy
|
± 0.2 °C Full Range
|
± 0.2 °C Full Range
|
± 0.35 °C at 50 °C ± 0.35 °C at 420 °C ± 0.5 °C at 660 °C
|
|
Stability
|
± 0.01 °C Full Range
|
± 0.02 °C at 33 °C ± 0.02 °C at 200 °C ± 0.03 °C at 350 °C
|
± 0.03 °C at 50 °C ± 0.05 °C at 420 °C ± 0.05 °C at 660 °C
|
|
Axial Uniformity at 40 mm (1.6 in)
|
± 0.05 °C Full Range
|
± 0.04 °C at 33 °C ± 0.1 °C at 200 °C ± 0.2 °C at 350 °C
|
± 0.05 °C at 50 °C ± 0.35 °C at 420 °C ± 0.5 °C at 660 °C
|
|
Radial Uniformity
|
± 0.01 °C Full Range
|
± 0.01 °C at 33 °C ± 0.015 °C at 200 °C ± 0.02 °C at 350 °C
|
± 0.02 °C at 50 °C ± 0.05 °C at 420 °C ± 0.10 °C at 660 °C
|
|
Loading Effect (with a 6.35 mm reference probe and three 6.35 mm probes)
|
± 0.006 °C Full Range
|
± 0.015 °C Full Range
|
± 0.015 °C at 50 °C ± 0.025 °C at 420 °C ± 0.035 °C at 660 °C
|
|
Hysteresis
|
0.025 °C
|
0.03 °C
|
0.1 °C
|
|
Operating Conditions
|
0 °C to 50 °C, 0 % to 90 % RH (non-condensing)
|
|
Environmental Conditions (for all specifications except temperature range)
|
13 °C to 33 °C
|
|
Immersion (Well) Depth
|
150 mm (5.9 in)
|
|
Insert OD
|
30 mm (1.18 in)
|
25.3 mm (1.00 in)
|
24.4 mm (0.96 in)
|
|
Heating Time
|
16 min: 23 °C to 140 °C 23 min: 23 °C to 150 °C 25 min: –25 °C to 150 °C
|
5 min: 33 °C to 350 °C
|
15 min: 50 °C to 660 °C
|
|
Cooling Time
|
15 min: 23 °C to –25 °C 25 min: 150 °C to –23 °C
|
32 min: 350 °C to 33 °C 14 min: 350 °C to 100 °C
|
35 min: 660 °C to 50 °C 25 min: 660 °C to 100 °C
|
|
Resolution
|
0.01 °
|
|
Display
|
LCD, °C or °F user-selectable
|
|
Size (H x W x D)
|
290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in)
|
|
Weight
|
8.16 kg (18 lb)
|
7.3 kg (16 lb)
|
7.7 kg (17 lb)
|
|
Power Requirements
|
100 V to 115 V (± 10 %) 50/60 Hz, 635 W 230 V (± 10 %) 50/60 Hz, 575 W
|
100 V to 115 V (± 10 %), 50/60 Hz, 1400 W 230 V (± 10 %), 50/60 Hz, 1800 W
|
|
Computer Interface
|
RS-232 and 9930 Interface-it control software included
|
|
|
|
Built-in Reference
Thermometer Readout Accuracy (4-Wire Reference
Probe)†
|
± 0.013 °C at -25 °C
± 0.015 °C at 0 °C ± 0.020 °C at 50 °C
± 0.025 °C at 150 °C ± 0.030 °C at 200 °C
± 0.040 °C at 350 °C ± 0.050 °C at 420 °C
± 0.070 °C at 660 °C
|
|
Reference
Resistance Range
|
0 ohms to 400 ohms
|
|
Reference
Resistance Accuracy‡
|
0 ohms to 42 ohms: ±0.0025
ohms 42 ohms to 400 ohms: ±60 ppm of
reading
|
|
Reference
Characterizations
|
ITS-90, CVD, IEC-751,
Resistance
|
|
Reference
Measurement Capability
|
4-wire
|
|
Reference Probe
Connection
|
6-pin Din with Infocon
Technology
|
|
Built-in RTD
Thermometer Readout Accuracy
|
NI-120: ± 0.015 °C at 0 °C
PT-100 (385): ± 0.02 °C at 0 °C PT-100
(3926): ± 0.02 °C at 0 °C PT-100 (JIS):
± 0.02 °C at 0 °C
|
|
RTD Resistance
Range
|
0 ohms to 400 ohms
|
|
RTD Resistance
Accuracy‡
|
0 ohms to 25 ohms: ±0.002
ohms 25 ohms to 400 ohms: ±80 ppm of
reading
|
|
RTD
Characterizations
|
PT-100 (385),(JIS),(3926),
NI-120, Resistance
|
|
RTD Measurement
Capability
|
4-wire RTD (2-,3-wire
RTD w\ Jumpers only)
|
|
RTD Connection
|
4 terminal input
|
|
Built-in TC
Thermometer Readout Accuracy
|
Type J: ± 0.7 °C at 660 °C
Type K: ± 0.8 °C at 660 °C Type T: ± 0.8 °C
at 400 °C Type E: ± 0.7 °C at 660 °C
Type R: ± 1.4 °C at 660 °C Type S: ± 1.5 °C
at 660 °C Type M: ± 0.6 °C at 660 °C
Type L: ± 0.7 °C at 660 °C Type U:
± 0.75 °C at 600 °C Type N: ± 0.9 °C at 660
°C Type C: ± 1.1 °C at 660 °C
|
|
TC Millivolt Range
|
–10 mV to 75 mV
|
|
Voltage Accuracy
|
0.025 % of reading + 0.01
mV
|
|
Internal Cold
Junction Compensation Accuracy
|
± 0.35 °C (ambient of 13 °C
to 33 °C)
|
|
TC Connection
|
Small connectors
|
|
Built-in mA
Readout Accuracy
|
0.02 % of reading +
0.002 mA
|
|
mA Range
|
Cal 4-22 mA, Spec 4-24 mA
|
|
mA Connection
|
2 terminal input
|
|
Loop Power
Function
|
24 V dc loop power
|
|
Built-in
Electronics Temperature Coefficient (0 °C to
13 °C, 33 °C to 50 °C)
|
± 0.005 % of range per °C
|
|
† The temperature range
may be limited by the reference probe connected
to the readout. The Built-In Reference
Thermometer Readout Accuracy does not include
the sensor probe accuracy. It does not include
the probe uncertainty or probe characterization
errors. ‡ Measurement accuracy
specifications apply within the operating range
and assume 4-wires for PRTs. With 3-wire RTDs
add 0.05 ohms to the measurement accuracy plus
the maximum possible difference between the
resistances of the lead wires.
|
Recently, we introduced Metrology Wells,
which provide the very best performance in portable temperature
calibration. The new 914X Series Field Metrology Wells extend
high performance to the industrial process environment by
maximizing portability, speed, and functionality with little
compromise to metrology performance.
Field Metrology Wells are packed with
functionality and are remarkably easy to use. They are
lightweight, small, and quick to reach temperature set points,
yet they are stable, uniform, and accurate. These industrial
temperature loop calibrators are perfect for performing
transmitter loop calibrations, comparison calibrations, or
simple checks of thermocouple sensors. With the “process”
option, there is no need to carry additional tools into the
field. This optional built-in two-channel readout reads
resistance, voltage, and 4–20 mA current with 24 volt loop
power. It also has on-board documentation. Combined, the three
models (9142, 9143, and 9144—each with a “process” option) cover
the wide range of –25 °C to 660 °C.
High performance for the
industrial environment
Field Metrology Wells are designed for the
industrial process environment. They weigh less than 8.2 kg (18
lb) and have a small footprint, which makes them easy to
transport. Optimized for speed, Field Metrology Wells cool to
–25 °C in 15 minutes and heat to 660 °C in 15 minutes.
Field environment conditions are typically
unstable, having wide temperature variations. Each Field
Metrology Well has a built-in gradient-temperature
compensation (patent pending) that adjusts control
characteristics to ensure stable performance in unstable
environments. In fact, all specifications are guaranteed over
the environmental range of 13 °C to 33 °C.
Built-in features to address large
workloads and common applications
Whether you need to calibrate 4-20 mA
transmitters or a simple thermostatic switch, a Field Metrology
Well is the right tool for the job. With three models covering
the range of –25 °C to 660 °C, this family of Metrology Wells
calibrates a wide range of sensor types. The optional process
version (models 914X-X-P) provides a built-in two-channel
thermometer readout that measures PRTs, RTDs, thermocouples, and
4-20 mA transmitters which includes the 24 V loop supply to
power the transmitter.
 |
| With the ability to maeasure a reference PRT, mA
current, and source 24 V loop power, Field Metrology
Wells can automate and save up to 20 different tests. |
Each process version accepts an ITS-90
reference PRT. The built-in readout accuracy ranges from
± 0.01 °C to ± 0.07 °C depending on the measured temperature.
Reference PRTs for Field Metrology Wells contain individual
calibration constants that reside in a memory chip located
inside the sensor housing, so sensors may be used
interchangeably. The second channel is user-selectable for 2-,
3-, or 4-wire RTDs, thermocouples, or 4-20 mA transmitters. For
comparison calibration, don’t hassle with carrying multiple
instruments to the field. Field Metrology Wells do it all as a
single instrument.
Traditionally, calibrations of temperature
transmitters have been performed on the measurement electronics,
while the sensor remained uncalibrated. Studies have shown,
however, that typically 75 % of the error in the transmitter
system (transmitter electronics and temperature sensor) is in
the sensing element. Thus, it becomes important to calibrate the
whole loop—both electronics and sensor.
The process option of Field Metrology Wells
makes transmitter loop calibrations easy. The transmitter sensor
is placed in the well with the reference PRT and the transmitter
electronics are connected to the front panel of the instrument.
With 24 V loop power, you are able to power and measure the
transmitter current while sourcing and measuring temperature in
the Field Metrology Well. This allows for the measurement of
as-found and as-left data in one self-contained calibration
tool.
All Field Metrology Wells allow for two
types of automated thermostatic switch test procedures—auto or
manual setup. Auto setup requires the entry of only the nominal
switch temperature. With this entry, it will run a 3-cycle
calibration procedure and provide final results for the dead
band temperature via the display. If you need to customize the
ramp rate or run additional cycles, the manual setup allows you
to program and run the procedure exactly how you would like.
Both methods are fast and easy and make testing temperature
switches a virtual joy!
Metrology performance for
high-accuracy measurements
Unlike traditional dry-wells, Field
Metrology Wells maximize speed and portability without
compromising the six key metrology performance criteria laid out
by the EA: accuracy, stability, axial (vertical) uniformity,
radial (well-to-well) uniformity, loading, and hysteresis. All
criteria are important in ensuring accurate measurements in all
calibration applications.
Field Metrology Well displays are
calibrated with high-quality traceable and accredited PRTs. Each
device (process and non-process versions) comes with an
IEC-17025 NVLAP-accredited calibration certificate, which is
backed by a robust uncertainty analysis that considers
temperature gradients, loading effects, and hysteresis. The 9142
and 9143 have a display accuracy of ± 0.2 °C over their full
range, and the 9144 display accuracy ranges from ± 0.35 °C at
420 °C to ± 0.5 °C at ± 660 °C. Each calibration is backed with
a 4:1 test uncertainty ratio.
New control technology guarantees excellent
performance in extreme environmental conditions. The 9142 is
stable to ± 0.01 °C over its full range and the mid-range 9143
is stable from ± 0.02 °C at 33 °C and ± 0.03 °C at 350 °C. Even
at 660 °C, the 9144 is stable to ± 0.05 °C. But this is not all!
Thermal block characteristics provide radial (well-to-well)
uniformity performance to ± 0.01 °C. Dual-zone control helps
these tools achieve axial uniformity to ± 0.05 °C at 40 mm (1.6
in).
Automation and documentation make
each unit a turnkey solution
 |
| The process version of Field Metrology Wells can
save up to 20 different tests. |
So you now have a precision calibration
instrument that has field-ready characteristics, accredited
metrology performance, built-in two-channel thermometry, and
automation—what else could you ask for? How about all this and a
turnkey solution that will automate and document the
results?
The process versions of Field Metrology
Wells have on-board non-volatile memory for documentation of up
to 20 tests. Each test can be given a unique alphanumeric ID and
will record block temperature, reference temperature, UUT
values, error, date, and time. Each test can be easily viewed
via the front panel or exported using Model 9930 Interface-it
software, which is included with each shipment. Interface-it
allows you to pull the raw data into a calibration report or an
ASCII file.
Operation is as easy as 1-2-3
You’ll find Field Metrology Wells
intuitive and easy to use. Each unit is equipped with a large,
easy-to-read LCD display, function keys, and menu navigation
buttons. Its “SET PT.” button makes it straightforward and
simple to set the block temperature. Each product has a
stability indicator that visually and audibly tells you the
Field Metrology Wells is stable to the selectable criteria. Each
unit offers pre-programmed calibration routines stored in memory
for easy recall, and all inputs are easily accessible via the
front panel of the instrument.
Never buy a temperature calibration tool
from a company that only dabbles in metrology (or doesn’t even
know the word). Metrology Wells from Fluke are designed and
manufactured by the same people who equip the calibration
laboratories of the world’s leading temperature scientists.
These are the people around the world who decide what a Kelvin
is! We know a thing or two more about temperature calibration
than the vast majority of the world’s dry-well suppliers. Yes,
they can connect a piece of metal to a heater and a control
sensor. But we invite you to compare all our specs against the
few that they publish. (And by the way, we meet our specs!)
|
|
|
Ordering Information for 9142
|
|
9142-X
|
Field Metrology Well, –25 °C to 150 °C, w/9142-INSX
|
|
9142-X-P
|
Field Metrology Well, –25 °C to 150 °C, w/9142-INSX, w/Process Electronics
|
|
X in the above model numbers to be replaced with A, B, C, D, E, or F as appropriate for the desired insert. See the inserts illustration and listing below.
|
|
9142-INSA
|
Insert “A” 9142, Imperial Misc Holes
|
|
9142-INSB
|
Insert “B” 9142, Imperial Comparison Holes
|
|
9142-INSC
|
Insert “C” 9142, 0.25-inch Holes
|
|
9142-INSD
|
Insert “D” 9142, Metric Comparison Holes
|
|
9142-INSE
|
Insert “E” 9142, Metric Misc Holes w/0.25-inch Hole
|
|
9142-INSF
|
Insert “F” 9142, Metric Comparison Misc Holes w/0.25-inch Hole
|
|
9142-INSZ
|
Insert “Z” 9142, Blank
|
|
|
|
|
Ordering Information for 9143
|
|
9143-X
|
Field Metrology Well, 33 °C to 350 °C, w/9143-INSX
|
|
9143-X-P
|
Field Metrology Well, 33 °C to 350 °C, w/9143-INSX, w/Process Electronics
|
|
X in the above model numbers to be replaced with A, B, C, D, E, or F as appropriate for the desired insert. See the inserts illustration and listing below.
|
|
9143-INSA
|
Insert “A” 9143, Imperial Misc Holes
|
|
9143-INSB
|
Insert “B” 9143, Imperial Comparison Holes
|
|
9143-INSC
|
Insert “C” 9143, 0.25-inch Holes
|
|
9143-INSD
|
Insert “D” 9143, Metric Comparison Holes
|
|
9143-INSE
|
Insert “E” 9143, Metric Misc Holes w/0.25-inch Hole
|
|
9143-INSF
|
Insert “F” 9143, Metric Comparison Misc Holes w/0.25-inch Hole
|
|
9143-INSZ
|
Insert “Z” 9143, Blank
|
|
|
|
|
Ordering Information for 9144
|
|
9144-X
|
Field Metrology Well, 50 °C to 660 °C, w/9144-INSX
|
|
9144-X-P
|
Field Metrology Well, 50 °C to 660 °C, w/9144-INSX,
w/Process Electronics
|
|
X in the above model numbers to be replaced with A, B, C, D, E, or F as appropriate for the desired insert. See the inserts illustration and listing below.
|
|
9144-INSA
|
Insert “A” 9144, Imperial Misc Holes
|
|
9144-INSB
|
Insert “B” 9144, Imperial Comparison Holes
|
|
9144-INSC
|
Insert “C” 9144, 0.25-inch Holes
|
|
9144-INSD
|
Insert “D” 9144, Metric Comparison Holes
|
|
9144-INSE
|
Insert “E” 9144, Metric Misc Holes w/0.25-inch Hole
|
|
9144-INSF
|
Insert “F” 9144, Metric Comparison Misc Holes w/0.25-inch Hole
|
|
9144-INSZ
|
Insert “Z” 9144, Blank
|
|
|
|
|
Ordering Information for All Field Metrology Wells
|
|
9142-CASE
|
Carrying Case, 9142-4 Field Metrology Wells
|
The following items are available for download in Adobe Acrobat (.pdf)
format and require Adobe Reader or Acrobat. Click on the link to view in
your browser or right click and “Save Target As...” to
download to your computer.
Note: Users Guides are available for download in
separate language versions. Visit our
manual page or go directly to our ftp site area -
ftp.hartscientific.com/manuals
Use our feedback form for more information.
|