Mini Fixed-Point Cell Furnace - 9260

9260

Features

  • Good introduction to fixed-point calibration
  • User friendly and inexpensive

Specifications 

Temperature Range  50 °C to 680 °C 
Ambient Operating Range  5 °C to 45 °C 
Stability  ±0.03 °C to 300 °C 
±0.05 °C above 300 °C 
Vertical Gradient  Top and bottom zones adjustable by offset 
Plateau Duration  6–10 hours typical 
Resolution  0.01 ° 
Display Scale   °C or  °F, switchable 
Immersion Depth  229 mm (9 in) 
Stabilization Time  15 minutes nominal 
Preheat Wells 
Fault Protection  Sensor burnout and short protection, over-temperature thermal cutout 
Display Accuracy  ±0.2 °C to 300 °C
±0.3 °C to 450 °C
±0.5 °C to 680 °C 
Comparison Block  Two multi-hole blocks, blanks, and custom blocks available 
Well-to-Well Gradient (in comparison block)  ±0.02 °C 
Heating Time  1.25 hrs. from 25 °C to 680 °C 
Cooling Time  10.5 hrs. from 680 °C to 100 °C 
Comm.  RS-232 included 
Power Requirements  115 VAC (±10 %), 60 Hz, 11 A, or
230 VAC (±10 %), 50 Hz, 6 A, specify, 1200 W 
Exterior Dimensions
(HxWxD) 
489 x 222 x 260 mm
(19.25 x 8.75 x 10.25 in) 
Weight  20.5 kg (45 lb.) with block 

Hart’s 9260 Mini Fixed-Point Cell Furnace provides a fixed-point system that cuts in half the financial investment required to do fixed-point calibrations and virtually all the time and training required by traditional systems. 

This furnace costs less than half of a large furnace and works with indium, tin, zinc, and aluminum cells to cover all ITS-90 fixed points from 156.5985 °C to 660.323 °C. The cells themselves, using a smaller volume of 99.9999 % pure metal, also cost much less. But cost is only a part of the issue. 

The 9260 makes using fixed points easy. Simply insert the cell at the end of the day and let it sit overnight. The next morning, initialize the built-in software routine for your specific cell. Come back in an hour, verify the stability of the cell, and you can take measurements for the rest of the day from a near-perfect temperature source! 

The built-in software lets you choose between using melting-point curves or freezing-point curves for each metal. The ITS-90 calls for freezing points, but melting points are easier to realize, and the difference in uncertainty (less than 2 mK for most applications) is generally insignificant. In fact, the difference between using traditional cells at their freezing points and Hart’s mini cells at their melting points is not significant for most labs in most applications. 

Comparison blocks are also available for the 9260 for high-precision comparison calibrations at high temperatures. Two blocks are available with a variety of pre-drilled wells in addition to blank or custom blocks. Well depth is 229 mm (9 in). 

Ordering Information

9260  Mini Fixed-Point Furnace 
SC1-FPMA Extended Warranty, Silver Care, FP Maintenance Apparatus
5914A  Mini Quartz Indium Cell 
5915A  Mini Quartz Tin Cell 
5916A  Mini Quartz Zinc Cell 
5917A  Mini Quartz Aluminum Cell 
5944  Metal Cased Mini Indium Cell 
5945  Metal Cased Mini Tin Cell 
5946  Metal Cased Mini Zinc Cell 
5947 Metal Cased Mini Aluminum Cell
2940-9260  Container, Mini-Cell Support, 9260 
2942-9260  Container, SST Mini-Cell Support, 9260 
1904-X  Accredited Cell Intercomparison 
3160-1  Comparison Insert, Blank 
3160-2  Comparison Insert, 7 holes at 6.35 mm (1/4 in) 
3160-3  Comparison Insert, 2 holes at 3.2 mm (1/8 in), 2 at 4.76 mm (3/16 in), 2 at 6.35 mm (1/4 in), 2 at 9 mm (9/32 in), and 2 at 9.5 mm (3/8 in) 
  Call for other comparison insert options. 