A temperature calibrator is used to remove inaccuracies in installed thermometers and temperature sensors, because temperature measurement is a critical parameter in many industrial processes. Although simulators are sometimes called temperature calibrators and may be used to provide calibration for the electronics in an installed system, a temperature source with a calibrated reference sensor is still required to provide a complete and meaningful calibration. Fluke’s Hart Scientific Division offers a complete selection of dry-block calibrators, Micro-Baths, and infrared calibrators (i.e. blackbody calibrators)for temperature calibration at the accuracies required by today’s industrial processes.
Micro-Baths
|
|
Model
|
Range
|
Accuracy
|
Description/Features
|
|
6102
Micro-Bath
|
35 °C to 200 °C
95 °F to 392 °F
|
±0.25 °C
|
Worlds smallest calibration bath.
Stability to ±0.02 °C.
Stirred 2.5-inch-diameter tank.
|
|
7102
Micro-Bath
|
5 °C to 125 °C
23 °F to 257 °F
|
±0.25 °C
|
Portable bath to 5 °C.
No refrigerationsolid-state cooling.
Stability to ±0.015 °C.
|
|
7103
Micro-Bath
|
30 °C to 125 °C
22 °F to 257 °F
|
±0.25 °C
|
Ultracold Micro-Bath reaches 30 °C.
No refrigeration or external cooling needed.
Stability to ±0.03 °C.
|
Handheld dry-wells
|
|
Model
|
Range
|
Accuracy
|
Description/Features
|
|
9100S
Handheld
Dry-Well
|
35 °C to 375 °C
95 °F to 707 °F
|
±0.25 °C at 100 °C
±0.5 °C at 375 °C
|
Worlds smallest dry-well.
Fixed block with 4-inch well depth.
Four hole patterns available.
|
|
9102S
Handheld
Dry-Well
|
10 °C to 122 °C
14 °F to 252 °F
|
±0.25 °C
|
Handheld unit cools to 10 °C.
Two 0.5-inch-diameter, removable sleeves.
|
Field dry-wells
|
|
Model
|
Range
|
Accuracy
|
Description/Features
|
|
9009
Dual-Block Calibrator
|
15 °C to 350 °C
5 °F to 662 °F
|
Cold block: ±0.2 °C
Hot block: ±0.6 °C
|
Dual-block industrial dry-well.
Each block has two wells with removable sleeves.
Water- and air-tight enclosure.
|
|
9103
Field Dry-Well
|
25 °C to 140 °C
13 °F to 284 °F
|
±0.25 °C
|
Small, lightweight field calibrator reaches 25 °C.
Stability to ±0.02 °C.
Calibrates up to six probes at once.
|
|
9140
Field Dry-Well
|
35 °C to 350 °C
95 °F to 662 °F
|
±0.5 °C
|
Portable field calibrator.
Choose from four multi-hole, removable inserts.
|
|
9141
Field Dry-Well
|
50 °C to 650 °C
122 °F to 1202 °F
|
±0.5 °C to 400 °C
±1 °C to 650 °C
|
High-temp field calibrator.
Interface-it software and RS-232 included.
Extremely small and fast for temperature range.
|
|
3125
Surface Calibrator
|
35 °C to 400 °C
95 °F to 752 °F
|
±0.5 °C to 200 °C
±1.0 °C to 400 °C
|
Calibrates surface sensors.
Plate stability of ±0.3 °C.
|
Infrared calibrators
|
NEW! Precision Infrared Calibrators
|
|
Model
|
Range
|
Accuracy
|
Description/Features
|
|
4180 |
–15 °C to 120 °C |
± 0.40 °C at –15 °C ± 0.40 °C at 0 °C ± 0.50 °C at 50 °C ± 0.50 °C at 100 °C ± 0.55 °C at 120 °C |
Fast, portable and easy to use
Correct target size for most thermometers
Calibration solutions from –15 °C to 500 °C (5 °F to 932 °F)
Radiometrically calibrated for traceable and consistent results |
|
4181 |
35 °C to 500 °C |
± 0.35 °C at 35 °C ± 0.50 °C at 100 °C ± 0.70 °C at 200 °C ± 1.20 °C at 350 °C ± 1.60 °C at 500 °C |
|
Model
|
Range
|
Accuracy
|
Description/Features
|
|
9132
|
50 °C to 500 °C
122 °F to 932 °F
|
±0.5 °C at 100 °C
±0.8 °C at 500 °C
|
Certifies most handheld pyrometers.
Short heating and cooling times.
|
|
9133
|
30 °C to 150 °C
22 °F to 302 °F
|
±0.4 °C
|
Calibrates at cold temperatures.
Gets to desired temperature quickly.
|
|
9135
3-Pt. IR Calibrator
|
50 °C, 100 °C, 150 °C
122 °F, 212 °F, 302 °F
|
±1 °C
|
Easiest to use IR calibrator.
Three set-pointsall reached quickly.
|
Metrology Wells
|
|
Model
|
Range
|
Accuracy
|
Description/Features
|
|
9170
|
45 °C to 140 °C
(49 °F to 284 °F)
|
±0.1 °C
|
Best-performing industrial heat sources (accuracy, stability, uniformity)
in the world.
Immersion depth to 203 mm (8 in).
Optional ITS-90 reference input reads PRTs to ±0.006 °C.
Temperature range from 45 °C to 700 °C.
|
|
9171
|
30 °C to 155 °C
(22 °F to 311 °F)
|
±0.1 °C
|
|
9172
|
35 °C to 425 °C
(95 °F to 797 °F)
|
±0.1 °C at 100 °C
±0.15 °C at 225 °C
±0.2 °C at 425 °C
|
|
9173
|
50 °C to 700 °C
(122 °F to 1292 °F)
|
±0.2 °C at 425 °C
±0.25 °C at 660 °C
|
Zero point dry-well
|
|
Model
|
Range
|
Stability
|
Description/Features
|
|
9101
|
0 °C
32 °F
|
±0.05 °C
|
Solid-state cooling.
Replaces messy ice bathseasy to operate.
Three wells, each 6 inches deep.
|
Dual block dry-well
|
|
Model
|
Range
|
Stability
|
Description/Features
|
|
9011
Hot Block
|
50 °C to 670 °C
122 °F to 1238 °F
|
±0.15 °C at 100 °C
±0.65 °C at 600 °C
|
Combined ranges from 30 °C to 670 °C.
1 unit 2 blocks.
Two independent temperature controllers (hot and cold side).
Stability: ±0.01 °C.
Multi-hole inserts hold up to 8 probes at once.
|
|
Cold Block
|
30 °C to 140 °C
22 °F to 284 °F
|
±0.25 °C (insert wells)
±0.65 °c (fixed wells)
|
Portable lab dry-wells
|
|
Model
|
Range
|
Stability
|
Description/Features
|
|
9007
|
40 °C to 140 °C
40 °F to 284 °F
|
±0.15 °C
|
40 °C with solid-state Peltier cooling.
Rugged truck-, plane-, and sea-worthy enclosure.
|
Furnaces
|
|
Model
|
Range
|
Stability
|
Description/Features
|
|
9150 Thermocouple Furnace
|
150 °C to 1200 °C
302 °F to 2192 °F
|
±0.5 °C
|
Benchtop thermocouple furnace.
Interchangeable insert sleeves.
Fast heating and cooling.
|
|
9112B Calibration Furnace
|
300 °C to 1100 °C
572 °F to 2012 °F
|
±0.1 °C
|
Standard block fits five probes.
Accommodates long probes.
Gradients less than ±0.3 °C at 1000 °C.
|