7080

Features

  • Self-contained refrigeration—no LN2 or chiller required
  • Temperatures as low as –100 °C in real metrology baths
  • Best stability and uniformity available at –60 °C and below
  • Large working areas for increased throughput

Specifications 

7060  7080  7100 

Range 

–60 °C to 110 °C 

–80 °C to 110 °C 

–100 °C to 110 °C 

Stability 

±0.0025 °C at –60 °C
 (methanol)
±0.002 °C at 0 °C
(methanol)
±0.0015 °C at 25 °C
(water)
±0.003 °C at 100 °C
(oil 5012) 

±0.0025 °C at –80 °C
 (methanol)
±0.0015 °C at 0 °C
 (methanol)
±0.0015 °C at 25 °C
(water)
±0.003 °C at 100 °C
(oil 5012) 

±0.008 °C at –100 °C
 (methanol) 

Uniformity 

±0.005 °C at –60 °C
 (methanol)
±0.005 °C at 0 °C
(methanol)
±0.003 °C at 25 °C
(water)
±0.005 °C at 100 °C
(oil 5012) 

±0.007 °C at –80 °C
 (methanol)
±0.005 °C at 0 °C
(methanol)
±0.003 °C at 25 °C
(water)
±0.005 °C at 100 °C
(oil 5012) 

±0.005 °C at –100 °C
 (methanol) 

Temperature Setting 

Digital display with push-button data entry 

Set-Point Resolution 

0.01 °C; high-resolution mode, 0.00007 °C 

Display Resolution 

0.01 °C 

Digital Setting Accuracy 

±1 °C 

Digital Setting Repeatability 

±0.01 °C 

Heaters 

500 and 1000 Watts 

350 and 700 Watts 

Access Opening 

127 x 254 mm (5 x 10 in) 

98 mm diameter (3.8 in) 

Depth 

305 mm (12 in) 

406 mm (16 in) 

Wetted Parts 

304 stainless steel 

Power 

230 VAC (±10 %), 50 or 60 Hz, 13 A, single phase, specify frequency 

230 VAC (±10 %), 50 or 60 Hz, 12 A, specify frequency 

Volume 

27 liters (7.2 gallons) 

18 liters (4.8 gallons) 

Weight 

159 kg (350 lb.) 

182 kg (400 lb.) 

Size (HxWxD) 

1168 x 775 x 483 mm
(46 x 30.5 x 19 in) 

1270 x 813 x 483 mm
(50 x 32 x 19 in) 

Automation Package 

Interface-it software and an RS-232 computer interface is included for setting the bath temperature via an external computer. For IEEE-488, add 2001-IEEE. 

Do you need a bath that chills below –40 °C to temperatures as low as –60 °C or even –100 °C? Would you like a bath that reaches those temperatures without using any external coolants? Hart has a variety of baths that meet these temperature requirements and give you the best stability in the industry. 

These baths are completely self-contained. They require no auxiliary cooling fluids or devices to achieve their set-point temperatures. Using Hart’s unique “heat-port” design, stability at –100 °C is ±0.0025 °C. No other company makes a bath that can match a Hart bath’s performance, and Hart baths are backed by our guarantee that if they don’t perform exactly the way we say they will, we’ll take them back. No arguments. No ifs, ands, or buts. 

Automate each of these baths with an interface package and Hart’s 9930 Interface-it software. If you want to completely automate the entire calibration process, see the description of Hart’s MET/TEMP II software package. 

Forget commodity-like utility baths! They’re not designed for high performance calibration needs. And be careful of companies that advertise performance specifications they don’t meet. It’s easy to write down numbers; it’s more difficult to meet them with an instrument. 

Remember, if our baths don’t perform the way we say they will, just send them back. Our equipment won’t disappoint you. 

Hart sells a complete selection of fluids for your bath. You can find these here.

Ordering Information

7060 

Standard Bath, –60 °C to 110 °C 

7080 

Standard Bath, –80 °C to 110 °C 

7100 

Standard Bath, –100 °C to 110 °C 

2001-IEEE 

IEEE-488

2010 

Access Cover, 127 x 254 mm (5 x 10 in), Lexan 

2007 

Access Cover, 127 x 254 mm (5 x 10 in), Stainless Steel 

2016-7060 

Fluid Level Adapter, 7060 

2016-7080 

Fluid Level Adapter, 7080 

2019-7100 

Fluid Level Adapter, 7100 

2069 

8X Magnifier Scope, with mounts 

2030 

Fast-Start Cooler 

 

Downloads

Right click one of the following links and “Save Target As ...” to download the catalog pages/data sheet or a product User's Guide in Acrobat (.pdf)  format.

(back to top)

Use our feedback form for more information.

 

 

 

Send mail to webmaster@hartscientific.com with questions or comments about this web site.
Copyright © 1998-2007 Hart Scientific, Inc.
Last Modified: 05/12/09