Buying the Right BathDuring a European trip we visited a lab struggling through the lab accreditation process. The hold-up was their bath. They had already tested baths from two manufacturers. The first bath didnt meet specs and the maker wouldnt rectify the situation, so the bath was returned. The second bath maker delivered a working bath, but when the accreditation auditor tested the bath he downgraded the labs accuracy class because they couldnt meet the required stability and uniformity levels. Most bath manufacturers tell you as little as possible about their baths performance. In fact, a few years ago one of our competitors used to tell people that high bath stability wasnt even necessary for accurate calibrations. Some still dont publish stability specs, and some are so elusive about the meaning of their specs that you can only conclude theyve got something to hide. Lab AccreditationAccreditation guidelines published by NVLAP specify that the temperature stability and uniformity of the bath fluid should be at least 10 times better than the required uncertainty of the sensor being calibrated. If youre testing a sensor with a modest specification of ±0.1°F over its whole range, your bath must be stable and uniform to ±0.01°F. Translated to Celsius, this figure becomes ±0.005°C, and you find yourself in need of a bath with performance to the third decimal place at each of the temperatures you must test. Several issues are involved in selecting a bath, and each item impacts calibration performance. StabilityStability is a measure of the baths control performance. How well does it maintain a constant temperature? Short-term instability is normally seen as an oscillation around the control point with its peaks defined in a ± statement. If the temperature of the bath fluid is changing during your measurements, you cant get reliable calibration results. Short-term stability is therefore absolutely crucial. Ask about short-term stability and define short-term as lasting at least 15 minutes. Less than that can prove very frustrating.
A baths stability will vary at different temperatures. Most baths perform best at temperatures close to ambient. The colder or hotter the set-point, the less stability. Too many sellers give you only one spec at or near ambient. Some give a single stability spec and dont ever mention that it applies only to one temperature or a narrow range. Ask about stability over the whole range that interests you. Bath fluid also affects stability. The higher a fluids viscosity and thermal constants, the larger the effect on stability. In addition to asking the temperature, ask what fluid was used when the spec was taken. For example, at 37°C a bath will be more stable with water as the medium. If youre going to use oil, expect somewhat larger instability. If your oil has high viscosity at 37°C, expect even greater degradation in stability. UniformityA bath can have good stability but poor uniformity. The bath must be homogenous in temperature throughout the test zone where youll make your comparison measurements. When you place two or more thermometers in the fluid, they should be at the same temperature during your measurement. The uniformity spec defines the peak value for this error source. The more probes youre testing, the larger the test zone and the more important uniformity becomes.
In a laminar flow bath (one where the fluid is stirred in a circular pattern), there may be no horizontal gradient, but because the fluid is not mixed vertically, there are gradients between different depths in the bath. This is a problem if your standards probe and the probes under test are not the same length. For example, if youre testing 3-inch-long probes and your standard is a 19-inch SPRT, youve got a problem. You can only immerse the test probes to 3 inches, but if you immerse the SPRT to only 3 inches you dont have sufficient depth to avoid stem effects and light piping that will affect the measurement made by the SPRT. If you properly immerse the SPRT and your bath suffers from vertical gradients, you wont be measuring the temperature at the 3-inch depth of your probes under test. Equilibration BlocksAccreditation guidelines recommend the use of a metal equilibration block to improve short-term stability during the measurement. However, a block can be inconvenient. The fixed location and diameter of its holes eliminate the flexibility of a bath to readily test any size or shape of thermometer. Youll need a new block for each probe type. Placing the probes in the block and the block in the bath is somewhat less convenient than simply dipping the probes directly in the liquid. Blocks also oxidize, and silicone oil will thicken and stick in the bottom of the holes. Regular cleaning is required to ensure continued performance levels. If youre testing many probes at a time, a block may not even work for you. It would be difficult to construct a block to properly test 20 thermometers at a time. Evaluate your bath purchase on specifications taken directly in the baths fluid. If youre given performance graphs, ask if a block was used. In your lab you can always add a block for the most critical measurements. Remember: the bath that performs the best without a block will also be the bath that performs the best with a block. Temperature RangeThe advertised temperature range of a bath is not necessarily the practical usable range. For example, a bath with a published range of -80°C to 150°C can be a bit misleading. The bath may operate over that temperature range, but currently theres no fluid to match that whole range. Those fluids that perform best at 80°C will evaporate too rapidly long before they get to 100°C, much less 150°C. An oil bath with an advertised range of 35°C to 300°C will be limited by the silicone oil you put in it. A good 300°C oil will be too viscous to deliver good performance below about 80°C, so with that fluid the baths range is 80°C to 300°C. In another example, a Hart salt bath works quite well at 40°C with the right fluid. But salt is molten only above 150°C. In addition to fluid, other factors mechanically limit a baths range. These include refrigeration, insulation, heater types and other material questions. Refrigeration gases break down above 150°C, thus limiting the life of the system. If a refrigerated bath is advertised with a higher range, ask if you must remove the cooling coil above a certain temperature. There are baths advertised with ranges from 80°C to 300°C in a single bath. However, the refrigeration gases or coils must be removed before going to the higher end of the temperature range. We could probably design a single bath that could operate from 100°C to 500°C. Besides the high price for such a bath, there would be no point. You would have to drain, clean and refill the bath at least three times during a calibration run in order to cover that range. The best solution to cover 100°C to 500°C is at least three baths with three different fluids. This way each bath design is optimized for performance in the range of the fluid you would use. Youll get the best stability and uniformity while tripling your throughput. Can You Ask Too Many Questions?Its not likely that a manufacturer will have a test file covering every temperature and fluid combination that interests you, but you can look for representative numbers. How many numbers will they give you? The more the better. If a salesman says his baths stability spec of ±0.005°C applies to the whole range, ask for a graph at several temperatures. If youre buying a bath for use at 300°C and the maker cant give you performance data above 100°C, you need to be skeptical. If a salesman talks about calibration accuracy instead of bath performance, ask for specific stability and uniformity data taken in the bath fluid. Finally, ask for a money-back guarantee of the performance. If you cant get what you need from the bath when its in your lab, will they take it back?
|
|
|
Click to request a quote, catalog, or other literature. Or call us at 1-800-438-4278. Home | What's New | Products | Support | U.S. Cal Lab | European Cal Lab | Accreditation |Training | Events | Publications |Contact Hart | International | Links | Fun Stuff | Software | ContentsSend mail to
webmaster@hartscientific.com with questions or comments about this web site. |
|